The world`s first plastic endodontic rotary
finishing file. This plastic rotary endodontic
finishing file has a unique file design and
diamond coating that enables the file to agitate
sodium hypochlorite and remove remaining dentinal
wall debris without further enlarging the canal.
The F® File Clinical Technique
1. The F® File should be the last file used
in a canal before obturation.
2. Prior to placement of the F® File, the canal
should be filled with sodium hypochlorite.
3. The F® File is inserted in an electric slow
speed handpiece that is set between 600-900
rpm.
4. The working length is set on the F® File
with a rubber stopper, then the file is placed
passively into the canal and circumferentially
worked along the dentinal walls with cyclic
axial motion (up and down).
5. The F® should be used in the canal for approximately
30 seconds, then the canal is flushed with sodium
hypochlorite and dried for obturation or placement
of an intracanal medicament.
6. The F® File is single usage per tooth and
should be properly discarded after use.
Additional Uses of the F® File
• Removal of calcium hydroxide dressing
• Placement of sealer
• Removal of canal debris after post preparation
• Placement of post cements
F® File Design & Specifications
• Size: 20
• Taper: 0.04mm
• Lengths Available: 21mm, 25mm, 31mm
• 6 pcs per box
• Radiopaque
• Diamond coated
• Single Use Only
Price: 10,50 €/box
|